Description
Perfectly grown single crystals have surfaces and sharp edges which are oriented to one another according to their main crystallographic planes. Light-optical goniometers are used
for determining the plane angles. In this process the angles of the plane normals are measured.
The goniometer 302 is designed for crystal dimensions of between 0,01 and approx. 5 mm. The positioning and read-out accuracies are 0,01° or mm respectively. This precision is
required for the preliminary adjustment on a goniometer head if the crystal is to be investigated by structural diffractometry.
Technical Data
| Horizontal Circle [°]: |
+/-140 |
| Vertical Circle unlimitted [°]: |
360 |
| read-out accuracy [°]: |
0,01 |
| Sample adjustment range [mm]: |
+/-10 |
| Read-out accuracy [mm]: |
0,01 |
| |
|
Microscope
Magnification 40-fold, 20-fold with additional lens
Ocular 10-fold, adjustable focus, with cross-hairs and fine scale. A scale unit corresponds to 0,025 mm on the sample (at 40-fold magnification)
Collimator
Websky cross-hairs
Double condensor optics
Projection lamp 6 V / 15 W
Adjustable illumination
Aperture disc with various apertures for limiting Websky cross-hairs